EMA 84 Instruction Manual106 Technical dataMeasurement datameasurement ranges 0...100 Pa0...1 kPa0...10 kPa0...100 kPaoverload capacity 10 x the final value of the measurement range(for measurement ranges 20 kPa)2 x the final value of the measurement range(for measurement ranges > 20 kPa)hysteresis 0.1 %time required for adjustment may toggle between 0.02 s, 0.2 s and 1 sdeviation from characteristic curvesetting limiting valuedeviation of reference: 0,3 Pa1 %0.5 % option (only for measurement ranges ≥100 Pa)0.2 % option (only for measurement ranges ≥1 kPa)Ambient conditionsmedium air, all non-aggressive gasesoperating temperature 0° C to +60° Cstorage temperature -10° C to +70° Crelative humidity 0…80 %EMC standards CEconformitydeclaration of conformity available upon requestElectrical datasupply voltage 9 V batteryoutput signal 0 to 1 V (BNC connector)load resistance RL RL 2 k for an output voltage of 0.. 1 Vdisplay 3½-place LCD, character height = 13 mmbattery life approx. 120 hours under constant operationPhysical datapressure port Ø 6.5 mm for NW5 tubing(interior tubing diameter = 5 mm)operating position preferably horizontaldimensions (w x h x d) 86 x 54 x 170 mmprotection class IP54weight approx. 0.75 kg, including batteryAppendix A: Parts in contact with measurement medium Beryllium bronze CuBe2 Araldite CY236 / HY988 Mu metal (nickel alloy) Loctite 242e Brass CuZn39Pb3 Carbonyl iron Aluminium AlCuMgPb / AlMg3 KEL (FPM: fluorinated rubber) Silicon (tubing), optional: Viton Vepuran Vu 4457/51 Crastin (PTBP) UHU-Plus endfest 300 binder