SECTION TWO - INSTALLATIONIDRONAUT – Brugherio (MB) OCEAN SEVEN 311 04-2019122. INSTALLATION AND START-UPUnpacking, installation and start-up procedures are described in this section.2.1. SHIPPING LISTDescription Quantity- OCEAN SEVEN 311 1- Dummy connector with looking sleeve 1- Laboratory RS232C communication cable 1x 3 m- 2x 1.5V Alkaline battery holder 1- 1x 3.6V Lithium battery holder 1- Sensor maintenance kit, including:pH reference sensor maintenance solution 1x 25 mlpH 7.0 buffer 1x 100 mlpH 4.0 buffer 1x 100 mlpH calibration cap 1- media kit, programs and documentation 12.2. INSTALLATION PROCEDUREUnpack and inspect the shipped parts; check the materials with respect to the above shipping list andbe sure that no damages have occurred during the transport. Open the probe and install the internalbatteries (see dedicated section). Remove the underwater dummy cap from the top cover of the probeand connect it with the connector of the laboratory cable.2.3. HOW TO OPERATE THE PROBEThe PROBE can be operated through the IDRONAUT Terminal Emulation software per Windows (forbrevity’s sake, hereafter referred to as “ITERM”). The “ITERM” software is described in a dedicatedOperator’s Manual. The communication parameters of the software must be set to meet those of theprobe (see the configuration sheet). The default communication values are: 38400 baud, 8 data bits, 1stop bit, no parity.After the installation, the “ITERM” programme must be configured to meet the default communicationset-up. Afterwards, any modification done to the OCEAN SEVEN 311 communication speed must bereflected on the ITERM programme too. ITERM initialization file keeps trace of the modification for thesuccessive start-up of the ITERM programme.Run the ITERM programme, set it up, and then switch on the probe by positioning the arm of therotating switch on the red dot. The OS311 now starts to wake up.NoteAt the start-up, the probe shows you the command prompt, waiting for the operator’s instructions, unless theprobe has been previously configured to perform self-recording data acquisition cycles. In this latter case, adedicated section of the manual describes how to revert the probe to normal operations.2.4. START-UP PROCEDURE2.4.1. Start-up - Verbose operating modeOnce the probe is switched on and the verbose operating mode is operative, the following messageappears:OCEAN SEVEN 311 - ID:311-0010101 {USR}[9.0_25-04/2019] 14:19:42.01 11-07-2019Work.Mem:Cnf[oK].Calb[oK].Stat[oK].WakeUp[9F]Data memory...[1 GByte]WarmUp-Battery..3.59 VDC....PROBE..Ok Analogue:.OkOperatingMode...Operator