English11Note: Connect tubing according to color code:Blue on Blue & Yellow on Yellow.4.4 Reusable Canister SystemQuick referenceThe assembly of the tubing isillustrated on the right side of theLSU.1 Connect the blue Aerosol filterto the blue Vacuum Inlet.2 Connect the yellow angledconnector to the yellowVacuum Seal.3 Connect the Patient Tubing tothe Patient inlet.4.5 After Use• When suction is complete, set OperatingKnob to ”0” and inspect all parts for damageor excessive wear. Replace parts if necessary.• Clean the LSU and its reusable partsaccording to section 6.2• Perform the Device Test according to section7.• Place the LSU on charge (see section 6.4).4.6 Decontamination of Laerdal ReusableCanister4.6.1 Parts to be DecontaminatedAfter Each Use:1 Canister Lid2 Float Ball3 Gasket.4 Vacuum Seal5 Angled Connectors6 Canister7 Patient Suction Tubing8 Suction Catheter Adapter.1.4.3. 8. 5.7.6.2.4.6.2 Preparation• Remove and empty the Canister accordingto section 4.3.• Disassemble the Canister into separate partsas indicated in the picture above.• The Float Ball on the reusable canister canbe snapped out of the lid.4.6.3 Rinsing• Rinse all parts under cold running water.• Immerse in warm water (30-40°C).4.6.4 Cleaning• Immerse all parts in hot water (60-70°C)containing a mild detergent (see section 6.3).• Thoroughly clean all surfaces, use a brushwhere possible.• Rinse in warm water and allow to dry.• Inspect all parts to be visibly clean and dry.Note: Thorough rinsing and cleaning are veryimportant steps prior to disinfection.