5Version ScreenSet Up MenuSetting OptionsLanguage: en / fr / de / it / da / es / fi /sv / no / ptLCD Contrast: 1 – 255 (Default 160)LCD Invert: Invert (✓) or non-invert ()Zero: Reset the stored Zero valueBeep: Enable (✓) or disable ()Active From: 2.0 to 40.0 % (default 2.0%)Zero ResettingZero resetting should only be performed if the TruCheck™ 2 Plus won’t zero in Track mode. If this is thecase the transducer may have been overstrained.If the reading is unstable the TruCheck™ 2 Plus should be returned to a supplier approved agent for repair. Itis recommended to get the TruCheck™ 2 Plus recalibrated as soon as possible after resetting the zero.Make sure no torque is being applied to the TruCheck™ 2 Plus and remove any tools from the input drivewhen resetting the zero.Active From SettingThis is the percentage of capacity at which Click mode and Dial mode will start to capture peak values.VersionP/N 43525S/N 124356S/W 1.0.1Press and hold from the measurement screento view the version screenThe version screen contains:P/N (Part Number of the TruCheck™2 Plus)S/N (Serial Number)S/W (Software Version Number)Press to return to the measurement screenorPress to enter the set up menuSet Up MenuLanguage enLCD Contrast 160LCD Invert ✓ZeroBeep ✓Active From 2.0%Select menu item (e.g. Language)Confirm itemChange setting (e.g. en to fr)Confirm settingSelect and press to save and exitCancel (exit without saving changes)