1 DO NOT measure or calibrate withoutthe protective probe guard in place.2 Immersion above the protective guardis not recommended. The cable can besubmerged briefly but is not designedfor continuous immersion.See “Probe Care and Maintenance”for more information.Immerseprobebeyonduppersteelband2 Getting Started8Press ON/OFF to power up your meter. Your meter will cycle throughvarious setup parameters when switched on.1 The first screen shows the model [Con6] [tdS6].2 The second screen shows the nominal cell constant value. The Micro 600COND and TDS meters can accept electrodes with k = 0.1, 1.0 or 10.0nominal cell constants. Default value is k = 1.0 [C 1.0]. See Section 5.2Advanced Setup to modify.3 The third screen shows the Normalisation Temperature which can be setto 25°C or 20°C. Default value is 25°C [t 25.0°C]. See Section 5.6Advanced Setup to modify.4 The fourth screen shows the Temperature Coefficient which can be set from0.0 to 3.0% per °C. Default value is 2.1%/°C [t 2.1%]. See Section 5.5Advanced Setup to modify.5 All LCD segments light for 2 seconds before entering measurement mode.Note: the meter will use the measurement mode that was inuse when it was previously powered off.2.8 Switching the Meter On