1720901330_EN_VVIEW (extd)_V1.11_230223Connection and operation44 Connection and operationConnection and operation4.14.1 ConnectionConnection4.1.14.1.1 InstallationInstallationNever install the measuring instrument too close to hot equip‑ment, like on top of a rotary vane pump, in order to avoid incor‑rect measurements or damages to the sensor.The gauge is intended for assembly directly to the apparatus(application).> Observe all specifications for installation, connection andoperation according to technical data, see chapter 9.1 Technical information9.1 Technical information.> Also observe rating plate data.> Compare the permitted limits which are described in thismanual, with your actual application regarding operatingmedia, pressures, forces, moments, temperatures and volt‑age.Installation conditions The gauge has acclimatized. Ambient conditions are observed and are within the limitationof use.Limitation of use (US)Ambient temperature 10–40 °C 50–104 °FAltitude, max. 2000 mabove sea level6.562 ftabove sea levelRelative humidity 30–85 %, non condensingProtection type/Impact energy IP 54/5 JPollution degree 2Avoid condensation or contamination by dust or liquids.Consider installationconditionsIMPORTANT!