Startup Operation48 FSM4000 D184B140U02Pos: 31.1 /Überschriften/1/G - I/Inbetriebnahme @ 0\mod_1131980668835_3101.doc @ 31506 Startup OperationPos: 31.2 /Überschriften/1.1/1-spaltig/J - L/Kontrolle vor der Inbetriebnahme @ 0\mod_1131981195011_3101.doc @ 34866.1 Preliminary checks prior to start-upPos: 31.3 /Inbetriebnahme/Durchfluss/FSM4000/Kontrollen vor Inbetriebnahme/Prüfpunkte (2008-03-20 10:06:26) @ 16\mod_1196081381078_3101.doc @ 142824The following points must be checked before commissioning:• The supply power must be switched off.• The supply power must match information on the model plate.• The pin assignment must correspond to the connection diagram.• Flowmeter sensor and transmitter must be grounded properly.• The temperature limits must be observed.• When flowmeter sensor (SE41F, SE21, SE21F) and transmitter (S4) are delivered as a pair,the data memory module (external FRAM) is located in the flowmeter sensor. The datamodule stores the flowmeter data, e.g., size, Cs, Cz, type, etc., as well as the setup data onthe transmitter after commissioning.• Prior to commissioning, plug the external FRAM for the appropriate flowmeter sensor (orderno. is printed on the FRAM and, if available, a TAG no.) to the connection board for theinstalled transmitter. Then screw to the connection board (captive).G0055811Fig. 471 External FRAMImportantIf a transmitter is ordered for an older model flowmeter sensor (see model number), anexternal FRAM is already connected to the connection board. You will also find the informationCs = 100% and Cz = 0%, which is required for flowmeters sensor from older product lines.See also chapter "11 Supplementary information: Operating S4 with an older sensor".