ProcessMaster FEP630, HygienicMaster FEH630 ELECTROMAGNETIC FLOWMETER | SI/FEP630/FEH630/FM/CSA-EN REV. A 19Connecting the signal cableOnly for remote mount design.The sensor housing and transmitter housing must be connectedto potential equalization.Terminal Function / commentsU FE Sensor power supplyGND GroundA Data lineB Data lineFunctional earth / ShieldingChange from two to one columnElectric data for operation in Div. 1Devices with HART protocolWhen operating in potentially explosive areas, observe the following electrical data for the signal inputs and outputs of thetransmitter.Current output terminals 31 / 32 / Uco can be operated on-site in active or passive mode through appropriate switching.Model: FEP631, FEH631 or FET632 Type of protectionOutputs on basic device ‘e’ / ‘XP’ ‘ia’ / ‘IS’UM[V]I M[A]UO UI[V]I O[mA]I I[mA]PO[mW]PI[mW]C O[nF]C I[nF]C OPA[nF]C IPA[nF]L O [mH] L I[mH]Current / HART output 31 / UCO, activeTerminals 31 / U CO30 0.2 30 30 115 115 815 815 10 10 5 5 0.08 0.08Current / HART output 31 / 32, passiveTerminals 31 / 3230 0.2 — 30 — 115 — 815 — 27 — 5 0.08 0.08Digital output 41 / 42, active*Terminals 41 / 42 and V1 / V2*30 0.1 27.8 30 119 30 826 225 20 20 29 29 0.22 0.22Digital output 41 / 42, passiveTerminals 41 / 4230 0.1 — 30 — 30 — 225 — 27 — 5 — 0.08Digital output 51 / 52, active*Terminals 51 / 52 and V1 / V2*30 0.1 27.8 30 119 30 826 225 20 20 29 29 0.22 0.22Digital output 51 / 52, passiveTerminals 51 / 5230 0.1 — 30 — 30 — 225 — 27 — 5 — 0.08All outputs are electrically isolated from each other and from the power supply.Digital outputs 41 / 42 and 51 / 52 are not electrically isolated from each other. Terminals 42 / 52 have the same potential.