OMNISTIM ® 500 PRO USER MANUAL - 25 -COPYRIGHT © 2004 - 2018, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVEDRange of Motion (ROM)The following protocol is found by pressing the ROM protocol button. The protocol displayed is MFACINCREASE ROM.A. MFAC INCREASE ROM:Provides e-stim-assisted passive or active assisted ROM at a joint. Target tissue is the superficial and/ordeep muscle groups. Bipolar set-up over agonist/antagonist muscles for targeted joint. Intensity should beset to elicit muscle contraction causing pain-free joint movement (typically a grade 3-4 muscle contraction).Activation of muscle causing contraction of the agonist and antagonist causes joint movement andimproved R.O.M.USER ADJUSTABLEPARAMETER (Y/N)ORDER INPARAMETERSET LOOPPARAMETER NAME ADJUSTABLERANGE (UNIT) DEFAULT (UNIT)Y 9 CARRIER FREQ 2.0, 2.5, 4.0, 5.0, 10.0 (kHz) 2.5 (kHz)Y 6 BURST FREQUENCY 0.0…250 (Hz) 35 (Hz)N BURST RATE SCAN 20 (%)N RATE SCAN ONN RATE SCAN TIME 4 (s)Y 1 TREATMENT TIME 0…99 (min) 15 (min)Y 2 8 2 B OUTPUT DELAY 0.0…9.9 (s) 0Y 4 RAMP UP 0.0…9.9 (s) 2.0 (s)Y 5 RAMP DOWN 0.0…9.9 (s) 2.0 (s)Y 2 T-ON 0…30 (s) 4 (s) 1Y 3 T-OFF 0…199 (s) 12 (s) 1Y 7 Channel Mode A/B SIMULTANEOUS, A/BALTERNATE, B DELAYED A/B ALT.1 When T-OFF = 0, T-ON = CONTINUOUS.2 B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.