17PHILIPS MEDIC A L SY ST EM STROUBLESHOOTINGThe following table provides troubleshooting tips.CPR Coachingventilations5 OFF: with ventilationsON: without ventilationsshock series 6 OFF: 1ON: 3protocol pause *(minutes)7 & 8 switch #7 switch #8 settingOFF OFF 2.0ON OFF 1.0OFF ON 1.5ON ON 3.0* If the Trainer is configured for CPR Coaching with ventilations, and the user selects CPR Coaching by pressing the i-button,the default pause and CPR Coaching settings (2 minutes pause, 30 compressions/2 ventilations) result in 5 cycles of CPRCoaching. The number of cycles varies for other settings. For all settings, CPR Coaching begins and ends withcompressions.† The CPR Coaching compressions/ventilations selection (switch 4) is applicable only if the CPR Coaching ventilations(switch 5) is set to OFF (with ventilations). Otherwise, the CPR Coaching compressions will be continuous, without breaksfor ventilations.PARAMETER SWITCH # SETTING (DEFAULT = OFF)BEHAVIOR POSSIBLE CAUSE RECOMMENDEDACTIONTrainer saysANALYZINGbefore pads areapplied tomanikin.• User touched theadhesive gel on thepads and activatedthe sensing circuitryof the Trainer.• Pad wires aretwisted.• Pad wires are notseparated.• Pads are touchingeach other.• Avoid touching theadhesive gel on thepads.• Untwist the pad wires.• Separate the pad wiresbetween the pads andthe red label.• Separate pads.