MODEL 6416/6417 SERVICE/CALIBRATION MANUAL CAT# 2141.01/2141.02v1 – 11/23/15 - 5 -LOOP INDUCTANCE MEASUREMENTMeasurement ranges (μH) Resolution (μH) Intrinsic uncertainty10 to 100 1 ±5% ±r100 to 500 1 ±3% ±rESTIMATE OF THE CONTACT VOLTAGEMeasurement range:- Contact voltage function: value calculated as the product of the loop impedance by theleakage current.Measurement ranges (V) Resolution (V) Intrinsic uncertainty0.1 to 4.9 0.1 ±5% ±r5.0 to 49.5 0.5 ±5% ±r50.0 to 75.0 1 ±10% ±rAlarm: range of thresholds from 1V to 75V.CURRENT MEASUREMENTMeasurement range:- Ammeter function: 0.2 to 40A. 4000-counts display.Measurement ranges (A) Resolution (A) Intrinsic uncertainty0.200 to 0.999mA 1μA ±2% ±50μA1.000 to 2.990mA3.00 to 9.99mA10μA ±2% ±50μA10.00 to 29.90mA30.0 to 99.9mA100μA ±2% ± r100.0 to 299.0mA0.300 to 0.990 A1mA ±2% ± r1,000 to 2,990 A3.00 to 39.99A10mA ±2% ± rAlarm: range of thresholds from 1mA to 40A.