MODEL 6416/6417 SERVICE/CALIBRATION MANUAL CAT# 2141.01/2141.02v1 – 11/23/15 - 6 -VARIATIONS IN THE DOMAIN OF USEThe influence is characterized by an Accuracy class number for each quantity of influence.Quantity ofinfluenceLimit of the Domain QuantitiesinfluencedInfluenceTypical MaxTemperature -20°C to +55°C A, Ω(1), Uc 1 P/10°C +r 2 P/10°C +rRelativehumidity10% RH to 90% RH A, Ω(1), Uc P+r 3 P+rBatteryvoltage4 to 6.5V A, Ω (1), Uc 0.1 P+r 0.25 P+rPosition ofconductorfrom the edge to thecenterA, Uc 0.1 P+r 0.2 P+rΩ (1) 0.05 P+r 0.1 P+rPosition ofclamp+/-90°, 180° A 0.2 P+r 0.4 P+rUc, Ω (1) 0.1 P+r 0.25 P+rProximity tomagneticmassSteel sheet 1mm thickagainst air gapA, Ω (1), Uc 0.1 P+r 0.5 P+rMagneticfield at 50 to60Hz30A/m A 2mA(2) 4.5mA(2)Uc 0.1 P+r 0,5 P+rFrequency ofthe current47 to 800Hz A, Uc P+r 2 P+rLeakagecurrent at50....60HzI<10ARxI<50VΩ (1) 2 P+r 8 P+r(1): Ω designates the quantities R, L, and Z.(2): Offset on the current measurement.POWER SUPPLY■ 4x1.5V LR6 (AA) alkaline batteries or 4 NiMH batteries.■ Mean consumption: approximately 140mA.■ Mean life: approximately 12 hours, or 14,400 30-second measurements.Remark: extreme environmental conditions may perturb the internal microprocessor. Simplydisconnecting the battery may be enough to eliminate this malfunction.The batteries should be removed for long-term storage.