Page 30 Aanderaa Data Instruments AS – TD302Figure 5-4 System configurationTo enable/disable a parameter:Select the output parameter from the list, press View/Edit in the lower part of the window, and change the settingby clicking the box (box is now checked), press Save to save and close the window.We recommend that you enable all parameters in case of later use. The memory card storage capacity is normallynot a limitation for the SeaGuard. Raw data can be used e.g. to control calibration coefficients and perform qualitycontrol on the data.The System Configuration holds output parameters which can be enabled/disabled by the user, refer Figure 5-4Select the property, and press View/Edit to set the value. Tap OK to return to the System Configuration display,refer Figure 5-4. Press Finish to store the settings when complete.5.3.2 Deployment SettingsThe Deployment Settings hold a list of user defined inputs: location, geographic position, vertical position andreference. To add information to one of these properties, select the property, press View/Edit and with thekeyboard panel enter the text or numbers. Press OK and Finish storing the information.5.3.3 User maintenanceNode Description, User Information and calibration coefficients are found in User MaintenanceOpen Administrative Tools from the Menu button Select User Maintenance and then Pressure Sensor from thesensor list. Refer Figure 5-5. In User Maintenance you find properties that are password protected, and areset/altered by the ‘advanced user’, note! the password is: 1000 The properties in user maintenance are thereforenot changed during normal operation. The user maintenance holds three submenus:• Node Descriptions• Owner• Calibration Coefficients