12 OMNISTIM® FX2 Pro USER MANUALCOPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVEDBattery Charger / Power Supply OperationBefore using the Omnistim® FX2 Pro battery charger/power supply: Verify that rechargeable batteries are installed within the Omnistim® FX2 Pro battery compartment. Check to see that the battery type selector switch located inside the battery compartment has been switchedto the rechargeable position.If you plug in the charger, after two seconds, the charge LED on the Omnistim® FX2 Pro comes on. After thebatteries are fully charged, the LED blinks. If you plug in the charger during treatment or when the unit is ON theLED turns off. (The charging continues but at a trickle and does not cause the LED to glow.) When you turn OFFthe unit after two seconds the LED comes on to announce charging.NOTE: If non-rechargeable batteries are used:1. Do not use the charger.2. Make sure that the rechargeable/non-rechargeable selector switch in the bottom of the batterycompartment is in the non-rechargeable position.Life expectancies of batteries under nominal (40mA MFAC) load conditions: 4.4Ah NiCad rechargeable batteries - 10 Hours. Alkaline non-rechargeable batteries - 100 Hours.Operational Sequence1. Press the POWER button (1/5) to turn on the power. The display panel will illuminate. There will be nostimulation current and the unit will be in the SELECT PROTOCOL / TREATMENT OFF mode.2. At this time select the treatment protocol desired and the appropriate treatment program by using the PROTOCOLbuttons (2) to select PAIN, NEURO RE-ED, MUSCLE RE-ED, FUNCTIONAL RE-ED, or FUNCTIONAL RE-EDASSISTED.3. Press the SELECT PARAM button to modify any settings. Press the MODIFY TIME/PARAM UP of DOWNbuttons to adjust the desired treatment time, or other settings, if you wish to change from the default settings.Set the timer for the desired treatment time. Pressing MODIFY TIME/PARAM UP or DOWN arrow buttonsincreases or decreases the duration from a minimum of 1 minute to a maximum of 60 minutes. The timer willcount down to zero during treatment. When zero is reached, a warning tone is emitted from the unit, thetreatment current is switched off and TREATMENT OFF is displayed. Pressing the START/STOP button (4)will shut off the warning tone. If the unit is left unattended, the warning tone is emitted for ten seconds beforepower is automatically switched off. Note that the treatment time is pre-set for each program but may beadjusted at will during treatment.4. Now position the electrodes on the patient, and connect the electrode plug(s). Adjust the output by pressing theOUTPUT A and OUTPUT B UP or DOWN buttons (3), to deliver appropriate output based on patientsensation. The screen will display the following (in this example, TENS-LVPC MOTOR was selected from theNEURO RE-ED protocol functional group.NEURO RE-EDTENS-LVPC MOTORSET OUTPUTS A-BA: 19 mA B: 19 mA 200 μS