24LimitationsThe Accu-Tell Blood Glucose Meter, Test Strips and Control Solution are beendesigned, tested and proven to work together effectively to provide accurate bloodglucose measurements. Do not use components from other brands.• Cat and Dog’s fresh capillary blood may be collected into test tubescontaining sodium heparin, lithium heparin if the blood is used within 10minutes. Do not use sodium fluoride/oxalate or other anticoagulants orpreservatives.• Use only with whole blood. Do not use with serum or plasma samples.• Very high (above 60%) and very low (below 30%) hematocrit levels cancause false results. Consult your veterinary doctor to find your pet’shematocrit level.• Abnormally high levels of vitamin C and other reducing substances willproduce false high blood glucose measurements.• The system is tested to accurately read the measurement of glucose inwhole blood within the range of 10 to 600 mg/dL (0.6-33.3 mmol/L).• Fatty substances (triglycerides up to 3,000 mg/dL (166.7 mmol/L) orcholesterol up to 500 mg/dL (27.7 mmol/L)) have no major effect on bloodglucose test results.• The Accu-Tell Blood Glucose Monitoring System has been tested andshown to work properly up to 10,000ft (3,048 meters).• Dispose of blood samples and materials carefully. Treat all blood samplesas if they are infectious materials. Follow proper precautions and obey alllocal regulations when disposing of materials.