Accutome AccuPach VI User’s Guidewww.accutome.com • (800) 979-2020 US & CA • (610) 889-0200 International • +44 (0) 1753-857177 Keeler Ltd.6Note: The power switch on the AccuPach VI device will not completely issolatethe device from power. In order to safely isolate the unit from power, theAccuPach VI must be unplugged.WARNING! BEFORE EACH PATIENT PROCEDURE, INSPECT THE PROBE TO ENSURE THATTHERE ARE NO BREAKS OR CRAKS IN THE OUTSIDE SHELL.WARNING! THE ACCUPACH VI PRODUCT IS ONLY TO BE USED WITH A MEDICAL-GRADEPOWER SUPPLY.WARNING! DO NOT POSITION THE ACCUPACH VI IN A WAY THAT WILL MAKE IT DIFFICULTTO OPERATE THE DICONNECTION DEVICE WHEN AN APPLIANCE COUPLER OR SEPARABLEPLUG IS USED AS ISOLATION MEANS.WARNING! THE ACCUPACH VI DEVICE CONTAINS A POTENTIAL EQUALIZATION CONDUCTORTHAT PROVIDES A CONNECTION BETWEEN EQUIPMENT AND THE POTENTIAL EQUILIZATIONBUSBAR OF THE ELECTRICAL INSTALLATION. THIS CONDUCTOR TERMINAL IS REQUIREDBY 60601-1:2005 TO BE ACCESSIBLE TO THE OPERATOR WITH THE ME EQUIPMENT INANY POSITION OF NORMAL USE AND SHALL ALLOW THE CONDUCTOR TO BE DETATCHEDWITHOUT THE USE OF A TOOL. THE RISK OF ACCIDENTAL DISCONNECTION SHALL BEMINIMIZED IN NORMAL USE.Avoiding Equipment DamageThe only peripheral equipment that may be connected to an AccuPach VI isa printer that has been selected and installed in compliance with EN60601-1requirements.The AccuPach VI provides no explosion protection from static discharge or arcingcomponents. Do not operate the instrument in the presence of explosive gasessuch as flammable mixtures of anesthetic and air, or nitrous oxide.WARNING! OPERATING OR STORING THE DEVICE BEYOND THE ENVIRONMENTAL RANGESIN THE SPECIFICATIONS CHAPTER MAY RESULT IN ERRONEOUS READINGS AND/ORPREMATURE FAILURE OR THE DEVICE.Avoiding Electromagnetic and Other InterferenceDo not use a cellular telephone and other devices not compliant with EMC Class Brequirements, as its signals may cause the equipment to malfunction. The effect ofradio signals on medical devices is dependent on various factors and thereforeunpredictable. To avoid electromagnetic interference, the device must beinstalled and operated in accordance with the user’s guide manual and using thecomponents supplied by Accutome.WARNING! THE USE OF ACCESSORIES, TRANSDUCERS, AND CABLES OTHER THANTHOSE SPECIFIED BY THE MANUFACTURER MAY RESULT IN INCREASED EMISSIONS ORDECREASED IMMUNITY OF THE ACCUPACH VI.WARNING! THE ACCUPACH VI SHOULD NOT BE USED ADJACENT TO OR STACKED WITHOTHER EQUIPMENT. IF ADJACENT OR STACKED USE IS NECESSARY, THE ACCUPACH VISHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN WHICHIT WILL BE USED.