(800) 979-2020 US & CA • (610) 889-0200 International • +44 (0) 1753-857177 Keeler Ltd. • www.accutome.com134. Select Meas and return to the Measurement screen. Table 3.1 belowprovides the IOP correction values.Table 3.1 - IOP Correction ValuesCorneal Thickness (micrometers) Correction Values (mmHg)405 7425 6445 5465 4485 3505 2525 1545 0565 -1585 -2605 -3625 -4645 -5665 -6685 -7705 -8Correction Values according to corneal thickness of 545 micrometers. Thesecorrection values are modified from the work of Doughty and Zamen. This chart wasreproduced from the Review of Ophthalmology, July 2002, Leon Herndon, MD, DukeUniversity, Glaucoma Service, Pages 88, 89, 90.How to Save a RecordAfter you have completed a patient exam,you may want to save the record to a USBflash drive. Using the flash drive allows theuser to save and print a patient’s recordto/from any computer.To save a record:1. Select the Util box on the LCD.2. Select the Save Ready box on theLCD.Note: A USB flash drive must be plugged into the USB port, otherwise “Save NoDisk” will be displayed. The USB flash drive can be plugged into the port at anytime to activate the Save Ready function.Figure 3.9 - USB Flash DriveGetting Started