26 User’s GuideThis monitor is intended for use in Health Care Facilities model AMM240WTDEquipment is not suitable for use in the presence of flammable anesthetic mixturewith air or with oxygen or nitrous oxide.No user serviceable parts inside, ask qualified personnel when accessing inside.For disposal of waste product, follow the requirement of local code.Electrical input rating: 24V DC 6.25AClassificationType of protection against electric shock: Class I Equipment.Degree of protection against the ingress of water: IPX1 compliance.Mode of operation: ContinuousThis monitor has been tested to comply with IEC/EN 60601-1, IEC/EN60601-1-2 andis certified by UL to medical standard UL60601-1(UL/cUL Mark).Because many medical offices are located in residential areas, this monitor, inaddition to the medical requirements, has also been tested and found to comply withthe limits for FCC Class B computing devices in a typically configured system. It isthe system integrator or configurer’s responsibility to test and ensure that the entiresystem complies with applicable EMC laws.Environmental conditions for transport and storage:- Temperature range within -4° to 140° F (-20° to 60° C)- Relative humidity range within 10% to 90%- Atmospheric pressure range within 500 to 1060 hPa.51LJMedical EquipmentE215822