USER MANUAL 2113.3 Drier working modes, drying time, sample timeDuring the balance – drier operation sampling of the mass on the pan takes place.Sampling time is set by the user, according to drying process speed. As a result ofsampling the current humidity value is calculated and displayed. Measurement is finisheddepending on selected Drying mode:1. In Time mode total humidity measurement time (Drying time) is defined by the user,2. In Short mode humidity measurement is finished, when drying is stopped anddifferences of a few successive mass samples are smaller than threshold value (2 mg).Amount of successive samples taken into consideration is defined as Samples quantity.Measurement is finished when Drying time is exceeded at the latest.>2mg<2mgDrying chart in Short mode for Samples quantity = 3.When choosing time mode only drying time and for example 10 times shorter sample timeis needed to start. In Short mode additionally Samples quantity is needed and Samplesinterval should be carefully calculated – end of drying is based on this parameter(and onSamples quantity).USER MENU1. Drying settings2. Memory settings3. Drying options4. Drying report5. Configuration6. Settings default7. ExitDRYING SETTINGS1. Drying temperature : 120C2. Mode : short mode3. Calculation : m0-m/m0*100%4. Samples quantity : 2 samples5. Sampling : 10s6. Drying time : 0:00:10s7. Drying profile : slow8. Settings storing : 1ExitmtQualifying amount(3 samples)StartStopTs