5-695-7-8. Configuration of CRITICAL MODEIntroductionCRITICAL MODE has the following setting and adjustment items:Item Contents ScreenROM VERDATEDisplays the ROM version anddate. 20.0 %> ROM VER.DATE YY-MM-DDSHIPPING DATA(default value)RECOVERYYou can return the device tofactory setting/default values forpertinent operational andconfiguration parameters.These parameters are enteredbefore the device is shipped, sothey are commonly referred to as“shipping data”.They include factory calibrationdata and factory settings or initialdefault settings for customerconfiguration data.20.0 %> SHIPPING DATARECOVERY READYINITIALDATARECOVERYInitial data recovery eliminatesall calibration data andconfiguration parameters.DO NOT use this function.20.0 %> INITIAL DATARECOVERY READYLCD display flowThe screen flow of CRITICAL MODE is as follows:> ROM VER. 3.00DATE 02-07-15> SHIPPING DATARECOVERY READY> INITIAL DATARECOVERY READY> OUTPUT CHECKMODE OFFCAUTIONCAUTIONINITIAL DATA RECOVERY function is only for Azbil Corporation service/maintenancespecialist. Please DO NOT use this function. If this function is turned ON, all calibrateddata will be missing. The device needs to be back to the factory to calibrate again.