Setup > Input/Output > Current #1Requires service level passcode or higher if security is enabled The current output, reset input and frequency/pulse/statusoutput can be set up through the SETUP > INPUT/OUTPUT menusAn asterisk (*) indicates the parameter defaultCurrent #1 Submenus Options/DescriptionsOUTPUT SOURCE*FLOW RATETEMPERATURE #1TEMPERATURE #2ENERGY FLOWVELOCITYSIGNAL STRENGTHTEST MODEDISABLEDSelect the reading to be assigned to the 4…20 mA outputTemperature and energy options only available with energy meterRANGE*4-20 mA4-20 mA NAMUR Current range is NAMUR 43 compliant with lower measuring limit at 3 8 mA and upperlimit at 20 5 mA and minimum alarm 3 5 mA and maximum alarm 22 6 mA0-20 mAMIN VALUE Enter the value of the reading at 4 mA Can also be the setting for the 0 mA setpoint when 4-20 mA RANGE isselected Units and decimal places based on parameter selected Negative numbers acceptedMAX VALUE Enter the value of the reading at 20 mA Units and decimal places based on parameter selected Negativenumbers acceptedFAILURE MODE*MIN CURRENTMAX CURRENTLAST VALUETEST CURRENTWhen an Fxx error occurs, such as low signal strength, the transmitter will setthe current output the selected valueFIXED VALUE Enter the value for the current output when there is a failure mode This parameter is only displayed withFAILURE MODE is set to FIXED VALUETEST CURRENT Available only when OUTPUT SOURCE is in TEST MODE Default 12 00 mA To check the wiring to the controlsystem or gauge, you can override the current output with a fixed current Numeric entry mA 0…22 mATRIM 4 mA Available only when OUTPUT SOURCE is in TEST MODE Set the test current to 4 mA or 0 mA, depending onthe current range selected Adjusts output until PLC/ DCS/BAS reads the desired valueTRIM 20 mA Available only when OUTPUT SOURCE is in TEST MODE Set the test current to 20 mA Adjusts output untilPLC/DCS/BAS reads 20 mAParameter Descriptions by MenuPage 32 October 2019TTM-UM-02222-EN-04