Installing the Control Panel in aRemote LocationFor your convenience, the control panel can be removedfrom the unit and mounted at a convenient location within10 ft. (3.1 m.) of the unit. To remove the control panelfrom the unit:1. Turn the unit OFF and disconnect it from thepower supply.2. Slide the control panel upward about 1/4” (.64cm) to 3/8” (.95 cm) - just enough to clear con-nector on top cover, pull forward, and remove.(The control panel fits tightly in the unit so it maybe difficult to remove.)3. Replace the control panel with the blank panelprovided with the unit to protect the electricalconnector.4. Using the wall bracket as a template, locate anddrill the mounting holes in the wall. A minimumof two (customer-supplied) fasteners will berequired. Attach the wall bracket to the wallusing the customer-supplied fasteners.5. Locate the 25-pin connector on the back topright of the unit. Remove the protective blackcover and store it for future use if the controlpanel is returned to the top cover.6. Locate the 10 ft. (3.1 m.) 25 pin M-F cableincluded with the unit; attach the female end tothe 25 pin connector on the back top right of theunit, and the other end to the connector on thebottom of the control panel.7. Tighten the cable screws on each end to securethe cable to the unit and control panel.8. The control panel may now be mounted on thewall.13INSTALLATIONNoteThe removable control panel was notdesigned to be repeatedly removedfrom the unit, therefore, you mayexperience difficulty when attemptingto remove it. It is thereforerecommended that the control panelbe permanently mounted in a remotelocation or remain intact with the unit.CautionRepeated removal and replacement ofthe control panel from the top covermay eventually cause it to becomedamaged.NoteThe blank protective cover on the rearside panel display connector shouldremain in place when the display isnot remotely located.BlankpanelWall MountBracket