124. Install the four screws removed in step 2through the bottom of the wall bracket andinto the EASYpure RO.Utility ConnectionsThe EASYpure RO is supplied with four separatelengths of tubing to be used for feedwater, rejectwater, feed to accessory storage reservoir and feedto another EASYpure unit's connections. Eachlength of tubing is six feet long and has the neces-sary fittings installed.Feedwater Connection1. Locate the length of 3/8" O.D. tubing pro-vided with a quick disconnect insert onone end and a 3/8" O.D. X 1/4 NPT tubingadapter on the other.2. Insert the quick disconnect insert into thefeedwater inlet on the EASYpure RO.(See Figure 3 for location.) Install the tub-ing adapter onto your incoming water line.We recommend a customer supplied shutoff valve be installed in your feed waterline.INSTALLATIONPower EntryModule/ON-OFF SwitchProductWaterOutletsPumpInterlockTank OverflowReject Water ConnectionFeedwater InletFigure 3 ConnectionsNoteWe recommend a customersupplied shut off valve beinstalled in your feed waterline.NoteThe quick disconnect fittingshave built-in shutoff valves inboth the coupling and theinsert.