33Fuse Replacement1. Disconnect the EASYpure RO from itspower source. Remove the outer cover.2. Locate the fuse holder, part of the powerentry module (see Figure 13 for location),and remove.3. Remove the burned out fuse(s) andreplace with one(s) of equal size and rat-ing.Membrane ReplacementA variety of factors will determine the service life ofyour EASYpure RO membranes. With proper main-tenance, the membranes should last approximatelyone year for normal usage.When either the "Service Membrane" or the"Replace Membrane" Indicator lights, use the fol-lowing procedure to replace the membranes:1. Disconnect the EASYpure RO from itspower source.2. Remove the membranes and discard.SERVICE AND REPAIRFuse HolderLocationFigure 13 Fuse Holder LocationWarningReplace fuses with those ofthe same type and rating.