MCDWIN ProgramComTec GmbH 5-11Lin / Log scaleFor a Lin scale all data intervals have the same size. With Log scale the intervals will be small forsmall y values and large for large y values. All options have effect only on the active Display.Calibration...Using the Calibration menu item or the corresponding icon opens the Calibration dialog box.Make a choice of several calibration formulas. Enter some cursor positions and the correspondingvalues, click on Add, then on Calibrate. The obtained coefficients can be inspected together withthe statistical error, or they can be changed and entered by hand. If ‘use calibration’ is on, thecalibrated values are displayed together with the channel position of the cursor.Comments...Up to eleven comment lines with each 60 characters can be entered using the Comments dialogbox. The content of these lines is saved in the data header file. The first line containsautomatically the time and date when a measurement was started. The titles of each line can bechanged by editing the file COMMENT.TXT.Figure 5.12: Calibration dialog box