Modifications Reserved 14OPM_ITS_USM_M06_M20_0US_V011 (09/2007)• When replacing batteries, replace with 12Volt/34Watt cell and exact quantity as listed below based on UPS VA rating:• 600VA – 2 batteries• 1000VA – 2 batteries• 1500VA – 3 batteries• 2000VA – 4 batteries• Please call your local GE representative for pricing and availability.9.5 Hot Swappable Battery ReplacementTower Procedure -Step 1. Remove front cover panel; pull out from top where finger grooves are located on each side. (See diagrams below)600 & 1000VA1500 & 2000VAStep 2. Push down slightly to allow front cover to be moved to the side. (See diagrams below)