B30 Patient Monitor3-22Document no. 2044677-001RESP - WAVE - NORMRATE - 20OHMS - 1.0RATIO - 1/1APNEA - OFFSHIFT - OFFCheck that the RESP waveform is shown and the RR value is 20 (±5). Change the positionof the BASELINE IMPEDANCE switch and check that appropriate RESP waveform and RRvalue are shown again within 30 seconds.RESP - APNEA - 32 SCheck that the monitor activates the APNEA alarm.NOTE: Make sure that only the ECG leads are connected to the simulator during theapnea test. If other cables are connected at the same time, the respiration signal from thesimulator may be disturbed, and therefore, the APNEA alarm may not be activated.NOTE: When you have the ECG service menu open, spikes will appear on the respirationwaveform. These spikes represent the threshold level for detecting inspiration andexpiration."Temperature measurement9. Communication and memoriesEnter the PSM: STP service menu:Parameters - STPCheck that the Time-outs, Bad checksums and Bad c-s by mod values do not increasefaster than by 5 per second. Check also that the STP board memories have passed theinternal memory test, i.e. the RAM, ROM and EEPROM show all OK."10. Temperature probe detectionCheck that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when noprobes are connected.Connect the temperature adapter cable to the module temperature connector and atemperature test plug to the adapter cable. Check that the Cable and Probe for T1 showON and the corresponding temperature value appears on the monitor screen."11. Calibration checkCheck the temperature calibrations using temperature test plugs.If the deviation on a temperature reading on the screen is more than 0.1°C, calibrate thetemperature channels according to the instructions in chapter “Temperature calibration”.”"12. Temp test