5010 Digital Communication InterfaceCommand DescriptionCommandFormatCommandExample ReturnedReturnedExampleAcceptableValuesDisplay TemperatureRead current set-point s[etpoint] s set: 999.99 {C or F} set: 90.00 CSet current set-point to n s[etpoint]=n s=80.00 InstrumentRangeRead temperature t[emperature] t t: 999.99 {C or F} t: 55.6 CRead temperature units u[nits] u u: x u: CC or FSet temperature units to Celsius u[nits]=c u=cSet temperature units toFahrenheitu[nits]=f u=fRead scan mode sc[an] sc scan: {ON or OFF} scan:ONSet scan mode sc[an]=on/off sc=on ON or OFFRead scan rate sr[ate] sr srat: 99.9 {C or F}/min srat:12.4C/minSet scan rate sr[ate]=n sr=1.1 .1 to 99.9Secondary MenuRead proportional band setting pr[opband] pr pb: 999.9 pb: 15.9Set proportional band to n pr[opband]=n pr=8.83 Depends onConfigurationRead cut-out setting cu[tout] cu cu:9999 {C or F} cu:105 C, inSet cut-out setting to n degrees cu[tout]=n cu-95 25 to 115°CRead heater power(duty cycle)po[wer] po po: 999.9 po: 1.0Configuration MenuRead high limit hl hl hl:999 hl:126Set high limit hl=n hl=90 25 to 100Read low limit ll ll ll:999 ll:-90Set low limit ll=n ll=-90 –90 to 25Read cooling setting co[ol] co cool: {ON or OFF} cool: ONSet cooling setting co[ol]=on/of[f] co=on ON or OFFRead serial sample setting sa[mple] sa sa: 9 sa: 1Set serial sampling setting to nsecondssa[mple]=n sa=0 0 to 999FULL or HALFSet serial duplex mode to full du[plex]=f[ull] du=fSet serial duplex mode to half du[plex]=h[alf] du=hTable 3 Controller Communications Commands