Q.seriesGantner Instruments Test & Measurement GmbH 574 Connecting the modules → Q.bloxx A106: Connecting sensors and I/Obridges the side drawn in dashes and the connection 10 are omit-ted.The bridge excitation voltage can be 2.5 V or 5 V (DC or CF).Fig. 4-42 A106, measurement with full and half bridges.TipInformation about the types of circuit and the respective advan-tages and disadvantages can be found in Section 6.3, Connectingsensors with sensing leads, Page 115.4.12.2 Strain-gauge quarter bridgesFor the connection of strain-gauge quarter bridges you need aspecial connecting plug which contains the completion resis-tances. The plug can be obtained under the designation Q.bloxxTerminal B14/120-A106 with120 or B14/350-A106 with 350 from Gantner Instruments Test & Measurement GmbH.The bridge excitation voltage can be 2.5 V or 5 V (DC or CF).IMPORTANTThe plug must have the same resistance values as the straingauges used, because otherwise no measurement is possible.Fig. 4-43 A106, measurement with strain-gauge quarter bridgeusing Q.bloxx Terminal B14.4.12.3 Inductive full and half bridgesWith inductive full bridges all connections are occupied. If thesensor has no sensing leads, you specify this during the moduleconfiguration (Type column). With half bridges the side drawn indashes and the connection 10 are omitted.The bridge excitation voltage can be 2.5 V or 5 V.9(10)5 (+)786 (–)120 Ω 350 Ω679679