Vers. No. 4.058 Released: 5/12/20114 Connecting the modules → Q.bloxx A106: Connecting sensors and I/OIMPORTANTDuring the configuration set CF 4800 Hz (carrier frequency) inthe Type column; inductive sensors cannot be operated withdirect voltage (DC) or 600 Hz carrier frequency.Fig. 4-44 A106, measurement with inductive full and half bridges.4.12.4 LVDT, RVDTWith (inductive) LVDTs or RVDTs all connections can be occu-pied, but you can also connect without sense leads and state thisin the Type column when configuring the module.The bridge excitation voltage can be 2.5 V or 5 V.IMPORTANTDuring the configuration set CF 4800 Hz (carrier frequency) inthe Type column; inductive sensors cannot be operated withdirect voltage (DC) or 600 Hz carrier frequency.Fig. 4-45 A106, measurement with LVDTs and RVDTs.4.12.5 Analog outputAn analog voltage output is available on each connecting plug.Fig. 4-46 A106, analog output.9(10)5 (+)6 (–)781095 (+)6 (–)78U3 (+)4 (GND)