ELECTROMAGNETIC ENVIRONMENT GUIDANCEELECTROMAGNETIC COMPATIBILITYNotice: The Gemini 810 + 980 diode laser complies with all requirements for electromagneticcompatibility according to IEC 60601-1-2: 2014.Description: The Activation Pedal uses Bluetooth BLE 4.0 technology, whichoperates at a frequency of 2402 to 2480 MHz with TX power of +0dBm and RXsensitivity of -93dBm and uses GFSK modulation. The pedal is pre-configuredby the manufacturer to only sync with the Gemini laser unit that has a matchingunique identifier. This prevents interference with other RF wireless technologiesthat may be present.As a safety measure, any termination of the Bluetooth link between the activationpedal and the laser unit during use will result in the immediate termination ofany laser emission. Reference the Service and Troubleshooting section of thismanual should you encounter any connectivity issues between the laser unitand the activation pedal.This device has passed wireless coexistence testing with common devices foundin dental practices at a minimum separation distance of 30 cm.Accessories: Medical grade power supply - Maximum length 6ft (1.8 meters)Ultradent P/N: 8981Activation Pedal: Wireless Bluetooth at 2.4GHzUltradent P/N: 8982 or 8992Medical electrical equipment needs special precautions regarding electromagneticcompatibility (EMC) and needs to be installed and put into service according to theEMC information provided in the following tables.Portable and mobile Radio Frequency (RF) communications equipment can affectmedical electrical equipment.The use of accessories other than those specified, except those supplied or sold byUltradent Products, Inc., as replacement parts for internal or external components,may result in increased EMMISSIONS or decreased IMMUNITY of the Gemini 810 +980 diode laser.CAUTIONWARNING44