1-28 0.5-mL Sample Block ModuleSetting Up the0.5-mL ProbeAssemblyTo set up the 0.5-mL Probe Assembly:Step Action1 If the heated cover is in the forward position, lift the lever, then slide the heatedcover back.2 Place the GeneAmp 0.5-mL Thermal Insulation Frame on the sample block.3 Use a cotton swab to coat the following wells with mineral oil:4 Place the 0.5-mL Probe Assembly into well A1.Note As the test progresses, you will move the 0.5-mL Probe Assembly to eachof the test wells.5 Thread the probe wire through the channel in the 0.5-mL Thermal Insulation Frameto prevent damage to the probe and lead wires.6 Make sure the probe is connected to the digital thermometer.7 Slide the heated cover forward and pull the lever down.IMPORTANT Seat the probe properly and close the heated cover carefully. Ifthe probe wire is crushed when the heated cover is closed, the probe may bedamaged.8 Turn on the digital thermometer.Note Refer to the instructions included with your Temperature Verification Kit for adetailed description on operating the digital thermometer.9 Continue with “Configuring the System 9700” on page 1-29.A1 D1A6 D7A10 E8B3 F1C5 F5C10 F10