1-32 0.5-mL Sample Block Module8 Slide the heated cover back and repeat steps 4–7 of “Setting Up the 0.5-mL ProbeAssembly” on page 1-28 and steps 2–7 of this procedure. Complete these steps forall 12 wells to be tested:9 The system 9700 evaluates the uniformity of the sample block temperature for thesetpoint values you entered and displays the results. A summary screen appears atthe conclusion of the test.If you entered values on the Temperature Non-Uniformity Test Data Sheet, comparethose values with the actual test results.10 Press Accept. To interpret the results, see “Evaluating the Results” on page 1-33.11 When you have completed all measurements, be sure to: Press Cancel. Remove the 0.5-mL Probe Assembly from the sample block. Turn off the digital thermometer and clean off the oil. Remove the 0.5-mL Thermal Insulation Frame from the sample block.IMPORTANT Make sure the sample block is at room temperature (~25 °C)before removing the frame.To run the Temperature Non-Uniformity Test: (continued)Step ActionA1 D1A6 D7A10 E8B3 F1C5 F5C10 F10WellF1 F3 F4 F5F2Accept Cancel94 °C 37 °CMorexx.xA1 xx.xxx.xA6 xx.xxx.xA10 xx.xxx.xB3 xx.xWell 94 °C 37 °Cxx.xC5 xx.xxx.xC10 xx.xxx.xD1 xx.xxx.xD7 xx.x