Option DescriptionCOLORIMETERHEATERSet the colorimeter heater settings from 20–60 °C (68–140 °F). The measuredvalue is shown.SAMPLE HEATER Set the sample heater settings from 20–60 °C (68–140 °F). The measured valueis shown.COLORIMETER Start an automatic test which increases the optical LED duty cycle in 5%increments. This starts from 0% until the output reaches the saturation. The A2Dcounts are shown for 0%, then % before the saturation and the first saturationvalue (%).STATUS LED Examine the front panel status LED indicator. The test cycles continuously untilinterrupted: off, red, green, yellow.A2D Set the colorimeter LED intensity to examine the cell transmittance for the A2Doutput.AIR PUMP Change and control the air pressure. SET SETPOINT—Range: 1–9.99 psi. LOWand HIGH DEADBAND—Range: 0–1 psi. SET LOW and HIGH VALUE—Range:5–99.99 psi. START—Start the air pump with the entered settings.FAN Set to adjust the fan duty cycle.ANALYZER TYPE For use by manufacturer technical support only.SELECT SCRIPT Toggle between normal instrument script and test script.SET CHANNELS For use by manufacturer technical support only.Replacement parts and accessoriesW A R N I N GPersonal injury hazard. Use of non-approved parts may cause personal injury, damage to theinstrument or equipment malfunction. The replacement parts in this section are approved by themanufacturer.Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer tothe company website for contact information.Replacement partsDescription Item no.Air pump filter 2718Bottle, reagent, 2 liter 9395000Capillary assembly, silica 6786900Capillary assembly, HR and LR phosphate 6786902Cell cap 6767800Cell shroud 6773100Colorimeter assembly, silica 6786800Colorimeter assembly, LR phosphate 6786801Colorimeter assembly, HR phosphate 6786802Colorimeter cell 6768000Colorimeter cover 6766900Fan assembly 6789800Fan filter plug 6789300English 15