Clean spillsC A U T I O NChemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional andnational regulations.1. Obey all facility safety protocols for spill control.2. Discard the waste according to applicable regulations.Sample line and valve cleanupC A U T I O NChemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protectiveequipment appropriate to the chemicals that are handled. Refer to the current safety data sheets(MSDS/SDS) for safety protocols.New tubing, valves and other sample conditioning equipment may be contaminated with silicate-based substances (oils, dust). These may contribute to slightly high readings until they are cleaned.1. Flush the sample line with sample for one to two hours.2. For an expedient procedure, inject one to four liters of a dilute caustic solution such as 1N (5%)sodium hydroxide solution into the front end of the sample line. Force the solution through theanalyzer to clean sample system components.Clean the sample cellC A U T I O NChemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protectiveequipment appropriate to the chemicals that are handled. Refer to the current safety data sheets(MSDS/SDS) for safety protocols.Put the analyzer in shutdown mode. Refer to Put the analyzer in shutdown mode on page 4.Clean the sample cell in the colorimeter as needed. Refer to Figure 1 and Figure 2.Items to collect:• Cotton swabs, wooden or paper. Do not use swabs with plastic sticks.English 5