Menu navigationPressing the “menu” button in the header bar calls the main menu. Thedisplay is made of three columns:• The left shows the menu options• The center shows a tree view of the position inside the menustructure• The right has the following generic controls:• Up - Return to previous menu (one step back)• Main - Jump directly to main menu• Close - Close the menu and go back to the measurement display.• Help - Help topics concerning current menuVirtual keyboardWhen a value or text is to be edited a virtual keyboard will appear on screen and can be used like astandard keyboard. Press CAP to access special keys. When input is complete press the Enter keyto confirm and exit the virtual keyboard. During editing, the edited field name is displayed along withunits where applicable.Security menuNote: When the instrument is started for the first time security is disabled. It is highly recommended that each userbe entered into the system and given appropriate access rights as soon as possible to avoid any unauthorizedaccess.Configure securityDefine access levels for all users . This requires a user access level 4.1. Select Configuration from the Security menu.Option DescriptionAccess rights When enabled only registered users can access the menus. When disabled (default)all menus are freely accessible and no ID is recorded against any action in the log file.Max session time The user is logged out automatically when the time limit is reached.User action logging When enabled every action from a logged on user is recorded in a user log file.User action log file The log file is a rolling buffer recording recent actions. Press Clear to empty the logfile.Access rights managementEach user has a unique ID and password used to:• Allow or deny a user to perform specific actions• To trace all actions by "ID" in a log fileOnce the ID and password are entered the user is allowed to perform actions according to the"Access level" that has been attributed by the Manager:Access level Typical rights0 View parameters, change views1 + Start / Stop measurements2 + Calibration14 English