D A N G E RExplosion hazard. Do not connect or disconnect electrical components or circuits to the equipmentunless power has been removed or the area is known to be non-hazardous.D A N G E RExplosion hazard. Connect only peripheral components that are clearly marked as certified for Class 1,Division 2 Hazardous Locations.Never connect any sensor or digital or analog module to an SC Controller that is not clearly markedas certified for Class 1, Division 2 Hazardous Locations.Product overviewThe 4–20 mA output module is an expansion card that supplies five 4–20 mA analog outputconnections to the SC4200c controller. The module connects to the expansion modules slots insidethe controller. The analog outputs are commonly used for analog signaling or to control otherexternal devices.Product componentsMake sure that all components have been received. Refer to Figure 1. If any items are missing ordamaged, contact the manufacturer or a sales representative immediately.Figure 1 Product components1 Module connector 3 Screwdriver, 2-mm wide blade2 4-20 mA output module 4 Label with wiring informationInstallationD A N G E RMultiple hazards. Only qualified personnel must conduct the tasks described in this section of thedocument.D A N G E RElectrocution hazard. Remove power from the instrument before this procedure is started.English 5