6. When the transfer is complete, push ENTER.7. To have the settings take effect immediately, restart the controller. Push the BACK key to exit theManage Configuration menu.Transfer settings to another deviceAll device settings including calibration, sensor name, selected temperature and measurement unitsand data logging settings are transferred.1. Push the MENU key.2. Select SD Card Setup and push ENTER.3. Select Manage Configuration and push ENTER.4. Select Transfer Settings and push ENTER. Two options appear:• Retrieve Settings• Copy Settings5. To retrieve settings from the controller (or a device connected to it) and put the settings on the SDcard:a. Select Retrieve Settings and push ENTER.b. Select the devices that contain the information to be transferred. All devices are selected bydefault. To deselect an item, highlight the selection and push the left arrow key. Push ENTERto begin the file transfers. Wait for the "Transfer complete" message.c. If files already exist on the SD card, a confirmation window appears. Select the devices againand push ENTER. Wait for the "Transfer complete" message.d. Push ENTER to return to the Manage Configuration menu.6. To copy settings from the SD card to a controller (or a device connected to it):a. Select Transfer Settings and push ENTER.b. Select Copy Settings and push ENTER.c. Select the devices on the SD card. All devices are selected by default. To deselect an item,highlight the selection and push the left arrow key. Push ENTER to begin the file transfers.7. When the transfer is complete, push ENTER to restart the connected devices.8. Push ENTER to restart the controller or push BACK to return to the Manage Configuration menu.Using the service portThe service port is used to download data files from the controller and install new versions ofcontroller and sensor firmware. To download data and update software, use the service port incombination with DataCom and a service cable (LZX887).Using DataComWhen using the service port, it is necessary to use DataCom. DataCom is a PC Application Utilitythat downloads data log and event log files from the controller and installed sensors. Files aredownloaded from the controller through the controller service port or they can be placed on a SecureDigital Memory (SD) card installed in the controller. In addition, DataCom is used to upload softwarefor the controller and sensors. The DataCom application must be installed on a PC to read the files.Refer to the DataCom manual for more information on how to use the application. The DataCommanual, software updates and other downloadable resources are available athttp://www.de.hach.com or http://www.hach.com on any sc200 product page.English 45