InstallationMounting components and dimensionsC A U T I O NPersonal injury hazard. Only qualified personnel should conduct the tasks described in this section of the manual.The controller can be installed on a surface, panel or pipe (horizontal or vertical). For mountingoptions and instructions, refer to Figure 2, Figure 3 on page 8, Figure 4 on page 9, Figure 5on page 10 and Figure 6 on page 11.For horizontal pipe mounts, the mounting feet (Figure 2) must be attached to the mounting bracket ina vertical position.For both horizontal and vertical pipe mounts, attach the mounting bracket to the controller as shownin Figure 5 on page 10.Figure 2 Mounting components1 Mounting foot (2x) 6 Flat washer, ¼-inch ID (4x)2 Sealing gasket for panel mount, Neoprene 7 Lock washer, ¼-inch ID (4x)3 Bracket for wall and pipe mounting 8 M5 x 0.8 Keps hexnut (4x)4 Vibration isolation gasket for pipe mount 9 Pan head screws, M5 x 0.8 x 100mm (4x) (Used forvariable diameter pipe mount installations)5 Vibration isolation washer for pipe mount (4x) 10 Pan head screws, M5 x 0.8 x 15 mm (4x)Note: A bracket for panel mounting is available as an optional accessory.English 7