Subject to change without notice 29according to IEC364-4-41 (CENELEC HD 384.4.41 and VDE 0100 Part410). Due to the circuitry of the HM6050-2 a maximum leakage currentof 800 mA can occur.It is not possible to operate the HM6050-2 at a mains outlet witha leakage current circuit breaker.Attention! Danger of death!The LISN has to be grounded according to the regulations (CenelecHD384/DIN VDE0100) separately; otherwise conductive parts of thehousing, which may be touched by the user, lie on a dangeroushigh voltage level.EMC measurement procedures for LISN HM6050-2Measurements have to be performed according to the VDE regulationsVDE 0877 Part 1.After the test setup is completed according to the regulations one canbegin with the measurement procedures.Follow these procedure steps:1. Set HM6050-2 to operation (power on)(Transient limiter in active status (LED off))2. Set DUT (Equipment under test) to operation3. Set Spectrum Analyzer/Test Receiver to operation4. Connect HM6050-2 to the input jack of the Spectrum Analyzer/TestReceiver via a coaxial HF cable.Visit us at www.TestEquipmentDepot.com