ConfigurationNormally, all instruments provided by ICT International come pre-configured,calibrated and tested. All that you need to do is select a logging interval. Thisis done from the Configuration screen:Note: Click ‘Update Measurement Option Changes’ after changing anysettings to send them to the logger.A range of standard Measurement Modes are available, from every 10minutes to every 60 minutes. In manual mode a measurement will be takenwhenever the Start Measurement button on the Channels screen is clicked.Cooling Time and Waiting TimeThe default settings are a cooling time of 10 seconds and a wait time of 6seconds. This ensures that water in the atmosphere of the chambercondenses and cools the thermocouple. A wait time of six seconds allows forstabilisation of the psychrometric wet bulb depression.