☞ Feed the steam pipe in until it stops.☞ Then lock the locking device by turning it clockwise by 60°(fig. 10).☞ Check the axial locking device on the steam pipe.Push-off mechanismThe plastic screw nut (E) helps loosen tight-fitting piston ground-inconnections (Fig. 10).☞ The plastic screw nut (E) helps loosen tight-fitting piston ground-in connections (Fig. 10).☞ To do so, hold the tight-fitting evaporator piston and turn the pla-stic screw nut (E) until the evaporator piston neck.Note: Prior to commissioning, hand-tighten the plastic screw nut(E) left-aligned. This will hold the steam pipe locking device (D)firmly in place.☞ Place the RV 06.15 or RV 06.13 condenser seal in the condenserreceptacle and fit the glassware to the device according to the as-sembly instructions (Fig. 11 a,b and c).Prior to commissioning - Mounting the seal RV 06.15☞ Insert vapour tube (1).☞ Insert seal RV 06.15 (2).☞ Mount the cooler (3a).☞ Tighten the cap nut hand-screwed. Start the device at 120 rpm for20 minutes. Then tighten the cap nut firmly (3b) .Assembling the vertical glassware cooler locking device☞ Assemble the condenser locking device according to the diagram(fig. 12).☞ Mount the plate (A) using the thumb screw (B).☞ Put the support rod (C) on the plate (B) and attach it with thescrew nut (D).☞ Attach the rubber protector (E).☞ Fasten the Velcro band (F) to the support rod (C).☞ Secure the vertical glassware with the Velcro band (F).Fig. 10EFig. 11bFig. 11a123a3bFig. 11cFig. 12BACDFE20 042009