6 AdjustmentThe balance must be adjusted at its place of installation before initial use and at regularintervals.A warm-up time of about 4 hours is recommended for stabilisation.Please use the recommended adjusting weight according to the technical data in theoperating instruction.6.1 Adjustment of non-verification balances6.2 Adjustment of verification balances (KERN 573-NM; KB-NM; FKB-M)Remark: The adjusting is only possible when not being blocked by the adjusting switch.The switch for this step is located at the bottom of the housing between the twoturning feet.Switch to the right Adjusting function is released.This setting has to be selected before starting the adjusting.Executing the adjusting according to below procedure.After the adjusting has been completed successfully it isnecessary to turn the adjusting switch to the left for blocking.Switch to the left Adjusting function is blocked.After the adjusting has been completed successfully theadjusting has to be blocked by switching to the left.The balance is now prepared for the procedure of verification.After the verification the adjusting switch has to be secured bya verification mark against access.572-SH-e-0951 7