7.2 Adjustment bulk weighing balance, not preconfigured by KERNOperation DisplayÖ Switch-on balance and during the self test press.Ö Use the number keys to enter password:Either• Standard password „0000“or• personal password, input see chapter 13Ö Confirm input with .Ö Select bulk material balance using .The appeared [T] displays the current balance.Acknowledge with .ÚÖ If necessary, at balance zero display usingselect the weighing unit [kg or lb], which shall beused for adjustment. The appeared [T] displays thecurrent weighing unit.Acknowledge with .Ö Ensure that there are no objects on the weighingplate. After successful stop check press the but-ton.Ö Enter value of the necessary adjustment weightusing numeric keyboard and acknowledge with .In order to achieve high-quality weighing results inthe sense of the measuring technology, it is recom-mended to select the nominal value as high as pos-sible. We recommend 80 % max.Ö Place the required adjusting weight carefully in thecentre of the weighing plate and press to con-firm.Ö After the adjustment the balance will carry out a self-test. Remove adjusting weight during self test,balance will return into weighing mode automatically.In case of an adjustment error or incorrect adjustingweight the display will show an error message( ), repeat adjustment process.CFS/CCS-BA-e-1013 19