FIS-BA-e-0512145 Calibration InformationWeight symbol flashing:Calibration switch (hardware seal) is not in secured condition.Calibration switch (=equalising plug)Traditionally, backup of the data relevant to calibration is carried out by means ofhardware backup (equalising plug on the circuit board in the terminal).Equalising plugon the circuit board in the terminalStatusPlug not inserted Calibration data not savedPlug inserted Calibration data savedThe calibration switch (equalising plug) can be seen through the right-hand windowon the terminal. This switch is secured by means of a sticker (stamp) when incalibrated condition.Balance calibration is not valid without the valid sticker.The four screws on the rear side of the terminal must be opened in order to reachthe calibration switch (equalising plug).Caution: Please remove the mains plug before opening the terminal.The GEO value can only be changed or the balance calibrated if the calibrationswitch (equalising plug) has been pulled out.