60 FTx/ITx-SH-e-0720Procedure• Ensure that the preload for which the calibration zero point is to be moved is in place.• Start the function. You will see a prompt asking you if you really want to move the calibration zero point. The calibrationzero point is moved when you confirm the prompt.If the display signals an underload or overload when you quit this menu, switch the scale off and then on again.7.13 Defining MinWeigh value(SCALE–> Minweigh)The MinWeigh function is to ensure that the weighing results lie within the defined tolerances corresponding to thespecifications of the quality-assurance system used.If the value for MinWeigh = 0, the menu for the supervisor is fully hidden.The value can be entered with the numerical keypad for KMT-TM terminals (ITT). Press the key for compact scales FTxand KMB-/KMS-TM terminals (ITB/ITS). The first digit starts flashing and can now be changed with the and keys.Confirm new digit with the key.7.14 Settings in the “TERMINAL” menu blockMost of the TERMINAL menu block is also available to the supervisor and is therefore not described here (see User Manual).However, this menu block includes a submenu that is restricted to the service technician and is described below.7.14.1 Defining a timeout (TERMINAL –> Timeout)In this menu you can limit the period of time for which the scale is used. This is particularly useful for leased scales, since itprevents the scale for being used for longer than the agreed period. This menu item is only available for KMT-TM terminals(ITT). The following settings are available:“Not active” There is no restriction on the period of use of the scale (factory setting).“Leasing” The duration of use is limited to the number of days entered. Enter the number of days (1 - 999)during which the scale can be used in the "Days" menu item. The duration of use is valid as of thenext day. If the set duration of use has passed, the scale still starts up, but no longer performsmeasurements and instead displays the message that the duration of use has passed.7.15 Setting of country versionAfter replacing the analog or digital board of the scale it is absolutely necessary to check the country specific settings and toadjust them if needed. The country specific settings affect the Technician Password, keyboard map, etc. Technician PasswordEU = , US = .For this procedure please connect the scale to an PC with an appropriated cable (PC – Scale, RS232, 9-pin straight cableSub D m/f) and open a terminal program on your PC (e.g. HyperTerminal). Verify that the scale is in “Dialog” Mode and thesettings of both interface of the scale and PC correspond.Depending on your location you may need to send the SICS command “DSK_US” (for USA and Canada) or “DSK_EU” (for allother countries) to the scale. The scale will send back “DSK_A”. This means that the command had been executedsuccessfully. If this is not the case, please re-check the connection between scale and PC as well as the interface settings andrepeat this procedure again. To inquire the current status of the “DSK”, send the SICS command “DSK”. Note that each SICScommand must be terminated with CRLF (carriage return, linefeed).