GAB/GAS/FAB-BA-e-0720 116.2 UnpackingCarefully remove the balance from the packaging, remove plastic cover and setupbalance at the intended workstation.6.2.1 SetupLevel balance with foot screws until the air bubble of the water balance is in theprescribed circle.6.2.2 Scope of deliverySerial accessories:KERN GAB / GAS / FAB• Balance• Operating Manual• Mains adapter6.3 Mains supply or battery operationPower is supplied via the external mains adapter. The stated voltage value must bethe same as the local voltage.Only use original KERN mains adapters. Using other makes requires consent byKERN.Output voltage mains adaptor: 3.2 V to 6.0 V CD (polarity: inside PLUS)The balance switches off automatically according to the pre-defined disconnectingtime. To achieve this, keep the MODE key pressed down for 6 seconds, then tap theMODE key repeatedly and store the selected disconnecting time by pressing theTARE key.The following applies:• P 0: Balance not turning off.• P 1: Balance turning off automatically after 20 minutes• P 2: Balance turning off automatically after 60 minutes