GAB/GAS/FAB-BA-e-0720 239 Instant helpIn case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Help:Fault Possible causeThe displayed weight does not glow. • The balance is not switched on.• The mains supply connection has beeninterrupted (mains cable not pluggedin/faulty).• Power supply interrupted.• Batteries are inserted incorrectly or empty• Maximum load for balance exceededThe displayed weight is permanentlychanging• Draught/air movement- • Table/floor vibrations• The weighing plate is in contact withforeign matter.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)The weighing value is obviously • The display of the balance is not at zerowrong • Adjustment is no longer correct.• Great fluctuations in temperature.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.