ITB-BA-e-072020Settings in the menuUser manual ME-Nr. 22017071 01/07Small platform model Large platform model1.Display "Scale":Press the -key (within 20sec)2.Display "Metrolo":Press the -key3.Display "ramp":Press the -key4.Display "SNR":Press the -key5.Display "SCAL.bld":Press the -key6.Display "GEO" (adjustment by means of GEO value):The GEO value can be set in this block and the balance also adapted to the local grav-ity ratios without adjusting weights.Case a)You are familiar with the GEO values.Once the -key has been pressed, the current GEO value will be displayed.Press the or -key, in order to alter the GEO value. . The next value is dis-played every time the key is pressed (adjustment range: 0 - 31). See the GEO valuechart in chapter 7.1.1 for the appropriate value.Confirm the selected GEO value using the -key.Please note: The GEO value may not be readjusted following this "adjustment bymeans of GEO value", as this would cause the set adjusting values to become invalid.Case b)You are NOT familiar with the GEO values. In this case adjustment must bemade using adjusting weights (see item 8).Press the -key.7.Display "LIN-CAL:" Press the -key.8.Display "CAL:" Press the -key.Press the -key. The scale determines the zero point. -preload- appears in the dis-play. The calibration weight to be placed on the scale then flashes in the display.If necessary, change the weight value displayed withPlace the calibration weight on the scale and confirm withThe scale calibrates with the calibration weight loaded. After calibration is completed,-done- appeats briefly in the display.