14 MWB-SH-e-1810.docx9 Error MessagesDisplay DescriptionZero range exceeded(on start-up or when pressing the key) Load on weighing pan Excess load, during zero setting of weighing scale Incorrect adjusting process Fault on load cellValue outside the A/D converter range Damaged weighing cell Damaged electronicsUnable to initialize zero point Measuring cell defective / overloaded Object on weighing pan / contact Transport safety device has not been removed Main board defectiveOL or -------- Overload Unload, switch off and adjust the balance------or „0“ Underload Unload, switch off and adjust the balanceShould other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.